ONTENTS
Code
Name
0B16
DSPH checksum error
0B17
DSPH self-diagnostics error
0B18
DSPH image processing error
DMA initialization error
0B20
/DMA device open error
0B20*
DMA device open error
0B21
DMA transfer error
0C00
DMA device open error
0C00*
DMA device open error
0C01
DMA transfer error
0C10
ID information read error
0C11
Image data read error
0C12
Image data read close error
*: A06 or later
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
DSPH program checksum error occurs.
An error occurs during DSPH program self-diagnostic.
Compression/decompression fails.
An attempt to open the DMA device fails.
An attempt to open the DMA device fails.
Image data transfer from the memory to the HCP board, or
from the HCP board to the memory, fails.
An attempt to open the DMA device fails.
An attempt to open the DMA device fails.
Data transfer from the memory to the CPU90F board fails.
An attempt to read the ID information from the HD to the
memory fails.
An attempt to read the image data from the HD to the
memory fails.
An attempt to close after reading the image data from the HD
to the memory fails.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
The HCP board is faulty, or DSPH program error.
The HCP board is faulty, or DSPH program error.
The HCP board is faulty, or DSPH program error.
The HCP board is faulty.
The HCP board is faulty.
The HCP board is faulty.
The CPU90E board is faulty.
The CPU90E board is faulty.
The CPU90E board is faulty.
HD failure.
HD failure.
HD failure.
MT - 21
Detail
Refer to
L-2
L-2
L-2
L-4
L-4
L-4
M-1
M-1
M-1
M-2
M-2
M-2
TR1H1235 .EPS
MT - 21