Replacement of MF1-related Timing Belt
(1) Loosen the screw that retains the tensioner.
(2) Remove the timing belt.
(3) Attach a new timing belt.
(4) Adjust the MF1-related timing belt.
" Adjustment of MF1-related Timing Belt"
Adjustment of MF1-related Timing Belt
(1) Loosen the screw that retains the tensioner.
(2) While checking the belt tension with a push-pull gauge, adjust the position of the
tensioner.
Adjust the position of the tensioner until a proper tension is attained, and then tighten
the screw to secure the tensioner in place. The tension is as follows.
500g, 6 ± 1mm
009-051-04
05.31.99 FM2460
3
4
5
6
Before-side-
positioning conveyor
Loosen screw
Before-side-
positioning conveyor
CR-IR341 Service Manual
7
8
9
10
MC1
(1)
Tensioner
Loosen
screw
(2)
Timing belt
FR1H1254.EPS
MC1
(1), (2)
Tensioner
Timing belt
FRONT
FRONT
500g
6–1mm
FR1H1252.EPS
MC - 96