FujiFilm CR-IR341 Service Manual page 279

Fuji computed radiography
Table of Contents

Advertisement

1
2
3
4
C
ONTENTS
Code
Name
0B00
Image data open error
0B01
ID information read error
0B02
ID information write error
0B03
Image data read error
0B04
Image data write error
0B05
Image data read close error
0B06
Image data write close error
0B10
DSPH device open error
0B11
DSPH program file open error
0B12
DSPH program read error
0B13
DSPH program boot error
0B14
DSPH memory write error
0B15
DSPH memory read error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
An attempt to reserve an image area of the HD fails.
An attempt to read the ID information from the HD to the
memory fails.
An attempt to write the ID information on the memory to the
HD fails.
An attempt to read the image data from the HD to the memory
fails.
An attempt to write the ID information on the memory to the
HD fails.
An attempt to close after reading the image data from the
HD fails.
An attempt to close after writing the image data on the
memory to the HD fails.
An attempt to open the DSP device fails.
An attempt to open the DSPH program file fails.
An attempt to read the DSPH program fails.
An attempt to boot the DSPH program fails.
DSPH memory write fails.
DSPH memory read fails.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
The HD is full.
HD failure.
HD failure.
HD failure.
HD failure.
HD failure.
HD failure.
The HCP board is faulty.
There is no dsphm1j.s24 file.
HD failure.
The HCP board is faulty.
The HCP board is faulty.
The HCP board is faulty.
MT - 20
Detail
Refer to
L-1
L-1
L-1
L-1
L-1
L-1
L-1
L-2
L-3
L-3
L-2
L-2
L-2
TR1H1234 .EPS
MT - 20

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents