FujiFilm CR-IR341 Service Manual page 186

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(2) Detailed Sequence of Search for Cassette Inlet IP
Fig. 5.3.2 Search for Cassette Inlet IP
Preprocessing: Fig. 5.3.1
Y
Suction cup
HP check
009-051-04
05.31.99 FM2460
3
4
5
6
START
Suction cup HP positioning
Fig. 2.3.2
Leak valve closed
DRIVE(SVB1/OFF)
Suction cup movement
(HP
load suction)
DRIVE(MB1/PB81)
Suction pump ON
DRIVE(PB1/ON)
Is SB5 closed within TB11?
N
Y
DELAY TB15
IP found (before search)
Suction leak
DRIVE(SVB1/ON)
Suction pump OFF
DRIVE(P1/OFF)
DELAY TB16
Suction cup movement
(load suction
HP)
DRIVE(MB1/PB82)
Is SB1 closed?
N
Level 2 error
23E8
Suction cup HP return error
Suction cup HP positioning
Fig. 2.3.2
Leak released
DRIVE(SB1/OFF)
END
CR-IR341 Service Manual
7
8
9
10
No IP
Table 5.3.2 Operation Parameters
Symbol
PB81
Suction cup movement (HP
PB82
Suction cup movement (load suction
TB11
Feed suction timeout
TB15
Wait for load suction
TB16
Wait for load leak
FR1H0406.EPS
MD - 123
Description
Value
load suction)
-14 p
HP)
14 p
1 sec
0 sec
0.1 sec
TR1H0059.EPS
MD - 123

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents