C
1
2
ONTENTS
(7) Position the following four types (five types if a mammo type is included) of cassettes
on arbitrary shelves.
IP conveyance starts.
• 14" x 14" (35 cm x 35 cm)
• 14" x 17" (35 cm x 43 cm)
• 10" x 12" (24 cm x 30 cm)
• 8" x 10" (18 cm x 24 cm)
• – (18 cm x 24 cm mammo)
(8) Conduct the following checks during IP conveyance.
• Check that the machine does not emit any abnormal sound or odor.
• Check that the IPs are conveyed in the order of cassette positioning.
• Check that the erasure lamp cooling fan (FANF1) operates.
(9) After completion of IP conveyance, remove all the cassettes to verify that the IPs are
not scratched.
(10) Change the positional relationships between the cassettes and shelves, and then
repeat steps (7) through (9).
The machine comes to an automatic stop when it repeats the IP conveyance
sequence 10 times.
(11) Remove the cassettes.
(12) Press the reset button.
The machine restarts and becomes ready for normal use.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
MT - 177
MT - 177