03DC After-Reading Conveyance Grip Operation Error
Cause
SE1 does not turn ON during after-reading conveyance grip HP detection.
Supplementary explanation
During IP conveyance, ME1 is driven in order to grip the after-reading conveyor.
When ME1 is driven, SE1 is turned ON/OFF by the actuator.
As SE1 turns ON, the HP is detected, resulting in the grip condition. As SE1 turns
OFF, the grip release condition is entered.
This error occurs when SE1 does not turn ON or OFF during the above operation.
Alternatively, this error occurs when SE1 fails to detect ON/OFF during initialization.