Replacement
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Power OFF the machine.
(2) Pull out the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(3) Disconnect the connector (CN1) of the LDD08A board.
(4) Remove the top cover from the scanning optics unit.
(5) Disconnect the connector (CN1) of the SYN08A board.
(6) Replace the SYN08A assembly.
(4)
Top cover
Screw x7
(3)
LDD08A CN1
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(6)
SYN08A assembly
BR M4x8
(5)
CN1
Spacer
Scanning optics unit
FR1H1176.EPS
MC - 159
MC - 159