FujiFilm CR-IR341 Service Manual page 502

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
To select an individual sensor to enable its monitoring
(1) Select "6. MECHANICAL UTILITY" and "3. SENSOR" in sequence.
(2) Select "2. MONITOR".
A list of sensor numbers is displayed.
(3) Type in the number of the sensor to be checked.
The machine is placed in a sensor monitoring enabled state.
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
> 6
0.QUIT
1.MOTOR
2.ACTUATOR
3.SENSOR
4.UNIT
MU > 3
0.QUIT
1.NUMBER
2.MONITOR
3.MONITOR ALL
MU:SEN > 2
MU:SEN > 2
1:SA1
6:SA6
11:SA11 12:SA12 13:SA13 14:SA14 15:SA15
16:SA16 17:SA17 18:SB1
21:SB4
26:SD2
31:SF3
INPUT THE NUMBER OF SENSOR
1 - 34 : 4
SA4 ----> Close
SA4 ----> Open
SA4 ----> Close
0
0.QUIT
1.NUMBER
2.MONITOR
3.MONITOR ALL
MU:SEN >
009-051-04
05.31.99 FM2460
3
4
5
6
2:SA2
3:SA3
4:SA4
5:SA5
7:SA7
8:SA8
9:SA9
10:SA10
19:SB2
20:SB3
22:SB5
23:SC1
24:SC2
25:SD1
27:SD3
28:SE1
29:SF1
30:SF2
32:SF4
33:SZ2
34:SZ3
CR-IR341 Service Manual
7
8
9
10
A
B
C
D
E
F
7
8
9
4
5
6
1
2
3
0
.
SP
DEL
BS
ENT
Caps
A
B
C
D
E
F
7
8
9
4
5
6
1
2
3
0
.
SP
DEL
BS
ENT
Caps
FR1H1235.EPS
MC - 11
MC - 11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus