(4) Loosely retain the positioning bracket by loosening its retaining screw.
(5) Loosely attach a new changeover guide drive solenoid assembly.
Be sure that the spring pin is lodged in the U-shaped groove in the arm.
(6) Install the spring.
(7) Connect the connector (CNSOLF1).
(8) Move the changeover guide drive solenoid assembly along the guide and then fully
tighten it.
Pull the changeover guide drive solenoid assembly to the left and above by sliding it
along the guide. When the arm begins to leave the stopper support, return the
changeover guide drive solenoid assembly until the arm comes into contact with the
stopper support, and then fully tighten the assembly.
SUPPLEMENTARY NOTE
Ensure that the arm is in contact with the stopper support but not depressed.
(9) Screw down the positioning bracket while pressing it against the changeover guide
drive solenoid assembly.
(5), (8)
Changeover guide
drive solenoid
Branch path changeover
009-051-04
05.31.99 FM2460
13
14
15
16
guide assembly
CR-IR341 Service Manual
(4), (9)
Positioning bracket
Guide
Spring pin
Arm
(6)
Spring
Stopper support
FR1H1413.EPS
MC - 242
MC - 242