(4) Insert a 300-mm rule along the right edge of the shelf.
Press the manual release lever to lower the hold pin.
(5) Secure the multi-stage cassette loading cover.
Loosely retain the two upper screws. After completing positioning steps (3) for the
uppermost shelf (first shelf), tighten the screws.
SUPPLEMENTARY NOTE
Because of parts machining precision, the multi-stage cassette loading cover may be slightly
warped when secured, though this will not pose any problem. Alignment with the cassette set
unit should take first priority.
(6) Connect connector CNT1.
(7) Reinstall the rest of the covers.
Reinstall the covers in the reverse order of removal.
" Removal"
(8) Make sure that the covers have been installed securely and there is no missing cover
screw.
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
Scale
(5)
Multi-stage
cassette loading
cover
(4)
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
(4)
Manual release
lever
Top view
(5)
0.5 - 1mm
(6)
CNT1
Scale
MC - 25
Cassette set unit
entry reference guide
(white plastic)
0.5 - 1mm
Multi-stage
cassette loading cover
FR1H1486.EPS
MC - 25