"10. SFD" (default value used when manual sensitivity setting is enabled)
The default value is switched when manual sensitivity setting is enabled in "9. SFC".
• 0: Standard
• 1: Semi-High-Speed
• 2: High-Speed
• 3: Menu Default (The content previously selected in the menu is stored in memory.)
"14. SS1" (screen saver timer)
The time until the screen saver is activated. Its initial value is "5" (minutes), and its
setting may be changed between 0 and 60 (in minutes).
When it is set to "0", the screen saver is disabled.
SUPPLEMENTARY NOTE
Screen saver: In order to protect burning of the operation panel screen, this function turns
OFF the screen after a certain period of time has elapsed with no user operation.
"17. ID_EDR" (EDR setup range)
The selection range for AUTO/SEMI/FIX/SEMI-X/MANUAL options of EDR function is
switched.
• 0: AUTO/SEMI/FIX can be selected.
• 1: AUTO/SEMI/FIX/SEMI-X can be selected.
• 2: AUTO/SEMI/FIX/SEMI-X/MANUAL can be selected.
"18. ID_FILING" (selection menu for image resolution during OD-F filing)
The menu for selecting the image resolution during OD-F filing is switched.
• 0: DISABLE (always saved in high resolution)
• 1: OFF/ON selectable
• 2: OFF/ON*/ON
ON*: The HQ image is reduced to the standard before transmission.
ON: The image is transmitted without reduction. For example, the HQ image is
transmitted as is.
"47. JSTFY" (definition of right or left justification for long format)
It is valid when "1. YES (long format used)" is set for "40. LTC" in "PRINT
(FILMFMT.CFG).
Here, a 16-digit variable value (0: left justified; 1: right justified) is set. Each of the 16
digits is assigned to "a" (patient name), "b" (patient name), "c" (exposure menu name),
"d" (examination number), "e" (department name), and "f"-"p" (reserved).
By default, the first five digits are "00010". Thus, "a" (patient ID), "b" (patient name),
and "c" (exposure menu name) are left justified; "d" (examination number) is right
justified; and "e" (department name) is left justified.