Adjustment
(1) Measure the tension value for the timing belt.
With a load of 300 ± 50g applied, the deflection should be 11 ± 1mm.
(2) Shift the tensioner position to adjust the tension of the timing belt until the specification
is met.
Erasure conveyor
009-051-04
05.31.99 FM2460
13
14
15
16
CR-IR341 Service Manual
Timing belt
FRONT
Tensioner
300g – 50g
11 – 1mm
MF1
MC - 238
FR1H1273.EPS