FujiFilm CR-IR341 Service Manual page 595

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(10) Remove the two screws and two spacers that retain the shaft of the before-side-
positioning retention guide, and take out the before-side-positioning retention guide
assembly.
(11) Remove the four screws and four spacers that retain the before-side-positioning
retention guide, and take out the before-side-positioning retention guide from the shaft.
Before-side-positioning
(12) Install new before-side-positioning retention guides to the shafts.
(13) Install the before-side-positioning retention guide assembly to the side-positioning
conveyor.
NOTES
• When installing the shafts, position their grooves on the right-hand side of the ma-
chine.
• When installing the before-side-positioning retention guide assembly, screw it down
while retaining the shafts with an Allen wrench lodged in the shaft hole.
(13)
Shock absorber-attached
before-side-positioning
(14) Reinstall the guide plate removed at step (2).
(15) Reinstall the before-side-positioning conveyor.
"6.1 Removing and Reinstalling Before-Side-Positioning Conveyor"
009-051-04
05.31.99 FM2460
3
4
5
6
conveyor
Shock absorber-attached
before-side-positioning
retention guide
Sems M3x10, x4
Allen wrench insertion hole
Flat machined
surface
retention guide
Groove
CR-IR341 Service Manual
7
8
9
10
(11)
Spacer x4
Shaft x2
(12)
Shock absorber-attached
before-side-positioning
retention guide
MC - 104
(10)
Shaft
BR M4x8, x2
Spacer x2
FR1H1553.EPS
FR1H1554.EPS
MC - 104