FujiFilm CR-IR341 Service Manual page 137

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.6
Barcode Reading and Side-Positioning
(1) Overall/General Sequence
This is for the operation of the side-positioning conveyor.
Barcode reading is performed in the middle of conveyance.
For 14"x17" and 14"x14" sizes, reverse conveyance is performed before side-positioning to
implement IP leading-edge positioning; for other sizes, this IP leading-edge positioning with
reverse conveyance is not necessary.
The side-positioning conveyance grip HP positioning is performed concurrently with the
after-reading grip HP positioning of the reading sequence.
For more detail, see 2.7 (3) and 2.7 (5).
Fig. 2.6.1 Barcode Reading and Side-Positioning (Overall Sequence)
In this sequence, the barcode
located on the back of the IP is
read during the IP conveyance.
For 14"x17" and 14"x14", reverse
conveyance is performed before
side-positioning to position the
leading edge of the IP.
IP side-positioning is performed.
Relationship between Conveyance Path and IP Driving
SF4
SF2
SF1
During barcode reading
009-051-04
05.31.99 FM2460
3
4
5
6
SB2
SF3
SC2
SC1
BCR
IP
SZ1SD3
CR-IR341 Service Manual
7
8
9
10
Preprocessing: Fig. 2.1.1
START
Does the machine comply with
IDT connection type?
N
Y
Barcode reading
Fig.2.6.2
Side-positioning
conveyance
Fig.2.6.3
Side-positioning
Fig.2.6.4
CSL/IDT information
check
Fig.2.6.6
END
SF4
SF3
SF2
SF1
SZ1SD3
Upon completion of
side-positioning conveyance
MD - 74
FR1H0331.EPS
SB2
SC2
SC1
BCR
IP
FR1H0332.EPS
MD - 74