To drive (DRIVE) FFM (MZ1)
(1) Select "6. MECHANICAL UTILITY" and "2. ACTUATOR" in sequence.
(2) Select "2. DRIVE".
A list of sensor numbers is displayed.
1:SOLA1
6:SVB1
11:FFM
INPUT THE NUMBER OF ACTUATOR.
1 - 11 :
Meaning: Enter the actuator number.
(3) Select "11. FFM".
The following message appears.
INPUT THE ROTATION DERECTION.
1.CW
Meaning: Enter the rotation direction.
X represents the rotation direction (either "1" or "2") that is currently set.
(4) Enter the direction rotation (either "1" or "2").
The following message appears.
INPUT THE SPEED.
1. STANDARD SUBSCANNING
2. RAPID SUBSCANNING
3. CARRY OUT
1 - 3 (FIXED=X) :
Meaning: Enter the rotation speed.
X represent the rotation speed (1: Standard subscanning; 2: Fast subscanning; 3:
Eject) that is currently set.
(5) Enter the rotation speed.
The FFM is driven.