FujiFilm CR-IR341 Service Manual page 349

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
03B2 Load Conveyance Error
Cause
After the reading is completed, the IP conveyed from the erasure conveyor does not
turn ON SB2. Or it remains ON, or it does turn ON but does not turn OFF.
Supplementary explanation
The IP conveyed from the erasure conveyor is conveyed to the up-down IP removal
unit by the driving of MC1/MB2.
The IP conveyed to the IP removal unit passes over SB2. (SB2: OFF to ON)
In order to return the conveyed IP back into the cassette, suction conveyance is
performed.
The IP is conveyed toward the cassette, and completely passes over SB2. (SB2: ON
to OFF)
This error occurs when the IP has been conveyed to the up-down IP removal unit but
fails to pass over SB2 (SB2 status does not transition from OFF to ON), or when it
does not completely pass over SB2 (SB2 status does not transition from ON to OFF),
during the above operation.
Sensor related
Error-related I/O locations
Cassette inlet IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SB2
SNS08A
CN
CN13
CN4
DRV03D
CN3
CN3
CNC2
DRV08A
CN1
CR-IR341 Service Manual
MB2
IP transport motor
MC1
IP transport motor
FR1H0011.EPS
CNB1
CNB2
CNMB2
CNB3
MB2
CNMC1
MC1
FR1H0012.EPS
MT - 80
MT - 80