FujiFilm CR-IR341 Service Manual page 464

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
4.4
Remedy When System Goes Down Unexpectedly during
Operation
Preliminary step 1:
1. Power OFF the machine.
Power OFF
Is the power supply normal?
Y
Preliminary step 2:
1. Remove the covers of the machine.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and, if any, remove it all.
Step 1
Are the boards set in the board
rack correctly?
Y
Step 2
Are the DIP switches and
jumpers set correctly?
Y
Step 3
Is the memory set in the
CPU90E board correctly?
Y
Replace the following boards.
MTH08A
CPU90E
CPU90F + ETH90F
DMC80A/HCP08A
IMG08A + IMG08B + IMG08C + IMG07B
SNS08A
DRV08A
SCN08A
JPS-1
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
Perform the recovery procedures described in 4.2 "Remedy When
N
Machine Does Not Power Up".
Then, restart the machine and check its status.
Set them correctly.
N
Then, restart the machine
and check its status.
Set them correctly.
N
Then, restart the machine
and check its status.
Set it correctly.
N
Then, restart the machine
and check its status.
CR-IR341 Service Manual
MT - 195
FR1H0140.EPS
MT - 195