FujiFilm CR-IR341 Service Manual page 560

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(4) Cut the cable tie.
(5) Disconnect the connectors (CNPB1).
(6) Remove the hose from the PB1.
(7) Remove the two screws (one of them should be merely loosened) that retain PB1, and
take out the PB1 from the pump assembly.
(8) Install a new PB1 to the pump assembly.
(9) Restore the cable tie cut at step (4).
(10) Attach the hose removed at step (6).
NOTE
When attaching the hose, connect it to the suction nozzle (upper one of the two
nozzles arranged) of the PB1.
Suction nozzle
Hose
(11) Connect the connector (CNPB1).
(12) Attach the bracket that was removed at step (3).
(13) Restore the cable clamp that was cut at step (2).
009-051-04
05.31.99 FM2460
3
4
5
6
IP removal unit
Timing belt
(7)
PB1
(7)
Remove
Sems M3x10
PB1
FR1H1496.EPS
CR-IR341 Service Manual
7
8
9
10
Up-down IP removal unit
(5)
CNPB1
MC - 69
(6)
Disconnect
(4)
hose
Cut cable tie
Pump
assembly
(7)
Loosen
Sems M3x10
FR1H1005.EPS
MC - 69