Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power ON
Does an error occurs when another
IP (IP with different dosage) is used?
Power ON
Is the sensitivity setting proper?
Replace the following parts in order named.
PMT board
SCN board
009-051-04
05.31.99 FM2460
3
4
5
6
N
Leave the error-causing IP alone for about 16 hours,
and then perform primary erasure before using it.
Y
N
From M-Utility, select "2. CONFIGURATION SETTING"
and "1. SYSTEM" in sequence to change the sensitivity setting
Y
(SFC/SFD).
CR-IR341 Service Manual
MT - 147
FR1H0119.EPS
MT - 147