Checking And Replacing Shock Absorbers Of Ip Removal Unit - FujiFilm CR-IR341 Service Manual

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.13 Checking and Replacing Shock Absorbers of IP Removal
Unit
Check and Replacement
Checking and replacing shock absorbers located under IP removal arm
(1) Rotate the timing belt to move up the IP removal unit to nearly the middle stage (away
about 150 mm from the tray of the cassette set unit).
(2) Check if the shock absorber is not peeled or frayed.
If the shock absorber is peeled or frayed, proceed to step (3) to replace the whole
guide to which the shock absorber is attached.
(3) Remove the spring from the wire.
(4) Remove the wire from the roller.
(5) Remove the left roller assembly from the shock absorber-attached guide plate.
(6) Remove the four screws (two of them should be merely loosened) that retain the right
roller assembly, and take out the roller assembly from the bracket.
(5)
Roller assembly
Sems M3x6, x2
(6)
Loosen screw
BR M4x8, x2
(6)
Roller assembly
(6)
Remove screw
BR M4x8, x2
(7) Pull out the IP removal arm toward the front.
NOTE
When pulling out the IP removal arm, do not grasp its center. If the center is grasped,
the IP removal arm may be bent.
009-051-04
05.31.99 FM2460
3
4
5
6
(2)
Shock absorber-attached
guide plate
CR-IR341 Service Manual
7
8
9
10
IP removal
(4)
Roller
(4)
Wire
(3)
Spring
(4)
Roller
Bracket
MC - 78
Up-down IP
removal unit
(1)
unit
Timing belt
(4)
Roller
FR1H1427.EPS
MC - 78