5.3
Checking, Replacing, and Adjusting Each Shelf Position
Sensor (SB3)
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
If anything abnormal is found during checks, replace or adjust SB3.
(1) Power ON the machine.
(2) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(3) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(4) Select "6. MECHANICAL UTILITY" and "3. SENSOR" in sequence.
(5) Select "2. MONITOR".
A list of sensor numbers is displayed.
(6) Type in the number of the each shelf position sensor (SB3).
The machine is now in the sensor monitoring enabled state.
(7) Make sure that the IP removal arm is accommodated.
If not, accommodate the IP removal arm.
(8) Rotate the MB3 timing belt counterclockwise, as seen from the rear, until the actuator
does not block the SB3 light path.
Visually check the actuator and SB3 through the hole located on the side plate of the
up-down IP removal assembly.
A message indicating that the sensor is OPEN appears on the operation panel.
(8)
Actuator
Hole
009-051-04
05.31.99 FM2460
3
4
5
6
SB3
CR-IR341 Service Manual
7
8
9
10
(8)
Timing belt
MC - 64
MB3
FR1H1241.EPS