Checking and Replacing Shock Absorber of Filter
(1) Remove the erasure conveyor.
"12.2 Removing and Reinstalling Erasure Conveyor"
(2) Remove the grip spring.
(3) Remove the two bracket retaining screws (one of them should be merely loosened)
and remove the bracket.
(4) Remove the lamp house assembly.
NOTE
When removing the lamp house assembly, use care not to damage the shock
absorber.
(5) Check if the shock absorber is not peeled or frayed.
If it is peeled or frayed, replace the whole filter to which the shock absorber is at-
tached. If it is not peeled or frayed, proceed to step (8).
(6) Remove the shock absorber-attached filter.
(3)
Remove screw
BR M4x8
Erasure conveyor
(7) Attach a new shock absorber-attached filter.
009-051-04
05.31.99 FM2460
13
14
15
16
(3)
Bracket
(6)
Only one shorter
screw is used.
CR-IR341 Service Manual
(2)
Grip spring
Erasure
conveyor
FR1H1624.EPS
(3)
Loosen screw
BR M4x8
(5)
Shock
absorber
Sems M4x6
MC - 244
(4)
Lamp house assembly
BR M4x8, x4
(6)
Shock absorber-attached filter
Spacer x5
BR M4x8, x5
FR1H1625.EPS