FujiFilm CR-IR341 Service Manual page 648

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(8) Disconnect the connector (CN1) of the LDD08A board.
(9) Remove the top cover from the scanning optics unit.
(10) Make sure that the connector (CN1) of the SYN08A board is locked.
If not, securely connect the connector to lock it in place.
Scanning optics unit
(11) Attach the top cover to the scanning optics unit.
(12) Connect the connector (CN1) of the LDD08A board.
(13) Push in the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(14) Make sure that the connector (CN2) of the SCN08A board is locked.
If not, securely connect the connector to lock it in place.
(15) Make sure that the no-fuse breakers (NFB6, NFB7, NFB9, NFB10) of the SCN08A
board are in their shorted conditions (button depressed).
Push the no-fuse breaker button several times to place the breaker in its shorted
condition.
NOTE
With board version J or later, the no-fuse breaker (NFB) is replaced with the fuse.
Thus, make sure that the fuse has not been blown.
(16) Power ON the machine.
Make sure that "HV-OFF" is displayed on the operation panel.
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
(4)
Upper cover
Screw x7
CR-IR341 Service Manual
7
8
9
10
MC - 157
(5)
SYN08A board
CN1
FR1H1175.EPS
MC - 157