[1] Checks on Image Trimming
CAUTION
Before adjusting the read start position and read width, set the image trimming to 0. If it is not
set to 0, the IP edge may be invisible; if so, the distance between the image frame and IP edge
cannot be measured.
(1) Activate M-Utility.
" Activating and Exiting M-Utility"
(2) Select "2. CONFIGURATION SETTING" and "2. PRINT" in sequence.
The FILMFMT.CFG file is opened to reveal its contents.
(3) Make sure that the image trimming (TE) is 0.
• If it is 0, touch the CANCEL button to take the steps described in "[2] Initialization of
Format Set Data."
• If it is not 0, write down the displayed value on your notepad, and touch the setup
value area.
(4) Set the image trimming (TE) to 0.
CAUTION
If the image trimming is changed by entering value 0, be sure to revert the trimming
back to its original value noted on your notepad after completing the adjustment of
read start position and read width.
(5) Touch the SAVE button.
The FILMFMT.CFG file is saved, and the screen returns to M-Utility.
(6) Take the steps described in "[2] Initialization of Format Set Data."
[2] Initialization of Format Set Data
Initialization of format set data should be performed when format-related setting has been
changed. It should also be performed when the scanning optics unit or any part in the scan-
ning optics unit is replaced.
(1) Activate M-Utility.
" Activating and Exiting M-Utility"
(2) Select "5. SCANNER UTILITY", "6. FORMAT", "1. DEFAULT" and "1. PIXEL AND
FREQ" in sequence.
The format set data is initialized.