1
2
C
ONTENTS
(6) Select "3. BACK UP", "3. NETWORK DATA" and "1.YES" in sequence.
> 7
0.QUIT
1.FORMAT FD
2.FORMAT IMAGE PARTITION
3.BACK UP
4.RESTORE
5.EDR DATA
6.PREVIOUS SYSTEM SOFTWARE
7.EXECUTION
FUT> 3
0.QUIT
1.SCANNER DATA
2.CONFIGURATION DATA
3.NETWORK DATA
4.IMAGE PROCESSING DATA
5.CSL MENU DATA
FUT:RES>3
Of the setup files, the following files are saved to the floppy disk.
• REMOTE SWITCH (RMT_SW.CFG)
• EQUIPMENT (EQUIP)
• NETWORK HOST INTERFACE (DEVICE)
• HOSTS ADDRESS (HOST)
• DISTRIBUTION (CODEDSTB)
• ROUTING (ROUTE)
• NETMASKS (NETMASKS)
• DICOM (DICOM)
NOTE
The menu parameters saved in the floppy disk also contain the language setting for the
machine. Thus, if the language setting differs during RESTORE and during BACK UP,
the file cannot be read during RESTORE.
(7) Select "0. QUIT" repeatedly.
The screen returns to the U-Utility mode.