FujiFilm CR-IR341 Service Manual page 671

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
10.4 Servicing Light-Collecting Mirror
CAUTIONS
• When removing and installing the unit, use care not to damage the cables.
• Wear gloves during checks and replacements. If not, the light-collecting mirror may be soiled
or damaged.
• Never touch the reflection face of the light-collecting mirror even when you wear gloves.
• Carefully handle the light-collecting mirror to avoid damage to it due to contact with surround-
ing objects.
• Wear a mask to prevent saliva from sputtering around on the light-collecting mirror.
Removal
(1) Remove the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(2) Remove the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
(3) Disconnect the connectors (CN1-3 from the SCN08A board).
(4) Remove the ground wire.
(5) Remove the four screws that retain the scanning optics unit.
(6) Carefully remove the scanning optics unit, using care not to shock it.
(7) Remove the two screws (one of them should be merely loosened) that retain the dust
cover, and remove the dust cover.
Slide the dust cover to the right (right-hand side of the machine), thereby taking out the
dust cover.
(7)
Dust cover
(7)
Remove screw
BR M4x8
009-051-04
05.31.99 FM2460
3
4
5
6
(7)
Loosen screw
BR M4x8
CR-IR341 Service Manual
7
8
9
10
(5), (6)
Scanner optics unit
BR M4x16, x4
MC - 180
(4)
Ground wire
BR M4x8
(3)
CN1—3
FR1H1810.EPS
MC - 180