Pushing In
(1) While using care to prevent the connector cables, etc. from getting caught, push the
tray of the cassette set unit all the way into the machine.
(2) Secure the tray of the cassette set unit with two screws.
(3) Connect the connectors (CNA1-3 to the cassette set unit; CNB1, 3, 4 to the up-down
IP removal unit).
(4) Attach the timing belt to the timing belt wheel.
(5) Put the tensioner back where it was.
Loosen the screw (BR M4x8) that retains the tensioner. Flick the timing belt several
times to ensure that it snugly fits onto the timing belt wheel, and then secure the
tensioner in place.
"3.4 Tensioner Not Requiring Tension Adjustment" in Machine Description
(1)
(2)
Tray of the cassette set unit
BR M4x8, x2
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
Up-down IP removal unit
Vertical conveyance unit
Cassette set unit
(3)
CNA3
CNA1
CNA2
MC - 27
(3)
CNB4
CNB3
CNB1
(4), (5)
Timing belt
BR M4x8
Tensioner
MC - 27