Replacement of Shutter
(1) Remove the shutter assembly.
" Removal of Shutter Assembly"
(2) Remove the bracket.
(3) Remove the torsion coil spring and slide bearings from the shutter.
(4) Attach to a new shutter the torsion coil spring and slide bearings that were removed at
step (2).
(2)
Bracket
Sems M3x6
(5) For shutter installation, reverse the removal procedure.
Installing the Shutter Assembly
Reinstall the shutter assembly.
For reinstallation, check the following points before screwing it down.
• The shutter assembly is pushed against both right and left side plates of the cassette
set unit.
• The cover of the shutter assembly opens and closes smoothly.