4.13 Replacing Guide Plate of Cassette Set Unit
(1) Remove the shutter assembly.
"4.3 Removing and Reinstalling Shutter Assembly"
(2) Disconnect the connector (SA3, 7, 11, 15) at the stage where the guide plate to be
removed is located.
(3) Remove the two grip springs of the rubber roller at the stage where the guide plate to
be removed is located.
(4) Remove the two bearings and two housings.
(5) Remove the rubber roller.
009-051-04
05.31.99 FM2460
3
4
5
6
(3)
Grip spring
(4)
Housing
(4)
Bearing
(5)
Rubber roller
CR-IR341 Service Manual
7
8
9
10
Cassette set unit
MC - 56
(4)
Bearing
(4)
Housing
(3)
Grip spring
FR1H1756.EPS