FujiFilm CR-IR341 Service Manual page 651

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(7) Reverse the removal steps to reinstall the SYN08A assembly.
NOTES
• When installing the SYN08A assembly, align the SYN08A assembly against the
reference face of the mounting bracket for proper positioning.
• The spacer that was attached before replacement should be used for reinstallation.
• Fit the cable into the cable guide.
Scanning optics unit
(8) Perform the procedures described in "Check after SYN08A Assembly Replacement."
009-051-04
05.31.99 FM2460
3
4
5
6
Top view
(8)
SYN08A assembly
CR-IR341 Service Manual
7
8
9
10
Mounting bracket
Reference face
FR1H1177.EPS
MC - 160
MC - 160

Advertisement