Adjustment of invalid pixels
(1) Activate M-Utility.
" Activating and Exiting M-Utility"
(2) Select "5. SCANNER UTILITY", "6. FORMAT", and "3. PIXEL ADJUST" in sequence.
(3) Enter the correction value.
Use the following input data. Note that 10 pixels correspond to 1mm in actual size on
the IP.
• To increment the current read start position by N pixels (0 < N ≤ 999)
Value of N: The image is shifted rightward relative to the image frame.
• To decrement the current read start position by N pixels (0 < N ≤ 999)
Value of -N: The image is shifted leftward relative to the image frame.
SUPPLEMENTARY NOTE
The MP (invalid pixel) value, N, must be set so that the value after correction ranges from 60
to 180. Thus, if the MP value is 150, up to 30 may be set for positive quantity of N, and up to
90 may be set for negative quantity of N.
To check the MP value, use "5. SCANNER UTILITY", "8. DATA MANAGEMENT" and "4.
DISPLAY DATA" menus.
If an out-of-range value is set, an error results. If an error occurs, check to see that the
procedures have been performed as instructed, and check if each of the parts is
connected properly.
009-051-05
11.20.99 FM2638
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
MC - 137
MC - 137