6.7
Replacing Cleaning Roller
(1) Remove the single screw that retains the cleaning roller assembly of the before-side-
positioning conveyor.
(2) Pull out the cleaning roller assembly horizontally.
(3) Remove the bracket.
Loosen the four screws located on right and left sides of the cleaning roller assembly
and remove the bracket.
(4) Remove the cleaning roller.
(4)
Cleaning roller
(5) Clean dust deposited within the cleaning roller assembly.
009-051-04
05.31.99 FM2460
3
4
5
6
(1)
Retaining screw
BR M4x8
(3)
Bracket
(3)
Loosen
sems M3x6, x4
CR-IR341 Service Manual
7
8
9
10
(2)
Before-side-
positioning
conveyor
(2)
Cleaning roller assembly
FR1H1441.EPS
FR1H1440.EPS
MC - 97
MC - 97