Replacement
(1) Remove the cassette set unit.
"4.2 Removing an Reinstalling Cassette Set Unit"
(2) Cut the cable tie.
(3) Disconnect the connectors (SA2, 6, 10, 14).
(4) Remove SA2, 6, 10, and 14.
Bracket
(4)
SA2, 6, 10, 14
Double sems M2x10, x2
(5) Install new SA2, 6, 10, and 14.
(6) Connect the connectors.
(7) Restore the cable tie that was cut at step (2).
(8) Adjust the sensors replaced.
" Adjustment"
009-051-04
05.31.99 FM2460