FujiFilm CR-IR341 Service Manual page 310

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
3910
SMCU transmission error
3911
Receive line break detection
3913
SMCU receive error
3920
SMCU transmission error
3921
Receive line break detection
3923
SMCU receive error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
A command transmission to the DMS fails.
A receive line break-ON is received.
A command reception from the DMS fails.
A transmission fails.
A receive line break-ON is detected while the operation is
stopped.
A command reception from the DMS fails.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
• Command wait timeout
• Break state on the target end
• Break state on the source end
• The serial line is disconnected.
• A failure occurs on the DMS.
• Command wait timeout
• Break state on the target end
• Break state on the source end
The DMC08A board is faulty.
• The serial line is disconnected.
• A failure occurs on the DMS.
• Command wait timeout
• Break state on the target end
• Break state on the source end
MT - 47
Detail
Refer to
E
F
E
E
F
E
TR1H1262.EPS
MT - 47

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents