Control Sheet Issue date Revision number Reason Pages affected 08.15.2005 New release (FM4693) All pages 11.10.2005 Corrections (FM4750) All pages CR-IR 357 Service Manual Safety Precaution CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
After the parts and retaining screws are restored to the above-mentioned states, follow the procedures set forth in this service manual to verify that the retaining screws are securely tightened to properly secure the parts. Safety-1 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 3
If the CR-IR 357 is installed in the patient environment with the connected equipment in the non-medically used room, the additional protective grounding will be required.
Page 4
fingers are not caught between the supporter - Always fix the machine by an optional anti-topple retainer. and machine. - Install the machine in a space with a height of the ceiling of 1750 mm or higher. Safety-3 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 5
"Laser Products - Conformance with IEC 60825-1, Am. 2 and IEC 60601-2-22; Final Guidance for Industry and FDA (Laser Notice No. 50)" and "EN60825-1 (Amendment 2).” issued by the FDA of the U.S. Safety-4 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 6
Safety-5 2.1.2 List of Laser Precaution Labels IEC60825-1:2001/EN60825-1:2002 Class 3B Panel Label HHS Certification and Identification Label Safety-5 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 8
Safety-7 Handling Instruction Labels and Attachment Locations 2.3.3 Side-Positioning Conveyor Unit 2.3.1 Cassette Set Unit 2.3.2 Erasure Unit Safety-7 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 9
Hold the portions of the green labels to dismount/mount the subscanning unit from/on the housing. 2.3.6 Scanning Optics Unit Hold the portions of the green labels to dismount/mount the scanning optics unit. Safety-8 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 10
Safety-9 2.3.7 Covers 2.3.8 Cassette Insertion Operation Label, Cassette Right-Justifying Label The cassette insertion operation label differs depending on the type of the cassette used in the user’s site. Safety-9 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 11
Safety-10 2.3.9 Lead Precaution Label 2.3.10 Power Cable Caution Label Apply the lead precaution label suitable for the local language. Power Cable Caution Label (Only Hospital Grade Power Cable for U.S.A.) Safety-10 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Page 12
flammable anesthetics mixture with air or with oxygen or nitrous oxide. Equipment not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or nitrous oxide. 5. According to the mode of operation CONTINUOUS OPERATION Safety-11 CR-IR 357 Service Manual 014-204-01E 11.10.2005 FM4750...
Guidance and manufacturer’s declaration - electromagnetic emissions The CR-IR 357 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-IR 357 should assure that it is used in such an environment. Electromagnetic environment -...
Page 14
The CR-IR 357 is intended for use in the electromagnetic environment specified below. The customer or the user of the CR-IR 357 should assure that it is used in such an environment. customer or the user of the CR-IR 357 should assure that it is used in such an environment.
CR-IR 357 The CR-IR 357 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CR-IR 357 can help prevent electromagnetic interference by maintaining a minimum distance...