Replacement of tensioner
(1) Remove the following covers.
Lower front cover, lower left-hand side cover
"3. Removing and Reinstalling Covers"
(2) Remove the transparent cover located on the left-hand side.
Using a blower, remove dust deposited on the transparent cover.
(3) Remove the spring.
(4) Remove the Kapton
(5) Replace the tensioner.
(3)
Spring
(6) For reinstallation, reverse the removal steps.
(7) Power ON the machine.
(8) Check MZ1.
Make sure that the motor rotates normally.
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(9) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04
05.31.99 FM2460