(10) Make sure that D13 located on the SCN08A board is lit.
• When D13 is lit:
Turn OFF the polygon and back ON again. If D13 is lit again, the polygonal mirror
assembly is normal. If it is not lit second time, it is conceivable that the polygonal
mirror assembly may be faulty. So, replace the polygonal mirror assembly.
• When D13 is not lit:
It is conceivable that the polygonal mirror assembly may be faulty. So, replace the
polygonal mirror assembly.
SCN08A board
009-051-04
05.31.99 FM2460
3
4
5
6
S1
D13
FR1H1277.EPS
CR-IR341 Service Manual
7
8
9
10
MC - 153
MC - 153