FujiFilm CR-IR341 Service Manual page 308

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
Code
Name
2D41
Compressed image cannot be accepted
2D43
HD full
2D44
Improper DICOM setup
2D45
Image data transmission error
2D50
File open error
2D51
File write error
2D52
File close error
User setup standard gamma parameter
3080
resetting
3717
SMCU transmission error
3718
Receive line break detection
3719
Serial line test error
371A
Serial line test error
371B
SMCU transmission error
*: "PDC" = DMA controller on the CPU90F board (PCI DMA controller)
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
5
6
Significance/Occurrence Condition
Transmission of a compressed image is refused by the target
device.
The HD of the output destination unit is full.
Negotiation for association with the output destination fails.
An error occurs during image data transmission.
The file cannot be opened.
An attempt to write a file fails.
An attempt to close a file fails.
Because the GP parameter is not defined when the
HIPUMN.PRM file is created during version update, the
default GP parameter is created (since only the MPF GP
parameter has been created).
Command transmission fails at the SMCU.
A receive line break-ON is received.
While waiting for a control line test response, a timeout
occurs.
While waiting for an operation permission command
response, a timeout occurs.
While waiting for a data block transfer response, a timeout
occurs.
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
The target device does not receive any compressed
image.
Free up the HD of the output destination unit.
• Based on DICOM information definition is incorrect.
• There is no setting for the own host on the output
destination.
• The transfer syntax or abstract syntax proposed
cannot be accepted by the output destination.
• DMA transfer error
• Timeout error
• Message queue generation error
• PDC driver device open error (*)
• No file
• HD failure
HD failure
HD failure
When parameters for any additional menu are present
in the user setup standard gamma parameter data.
The SMCU driver is faulty.
• The serial line is disconnected.
• A failure occurs on the IDT.
• A failure occurs on the IDT.
• The SMCU driver is faulty.
• A failure occurs on the IDT.
• The SMCU driver is faulty.
• A failure occurs on the IDT.
• The SMCU driver is faulty.
MT - 45
Detail
Refer to
G-2
G-2
G-2
G-2
A
A
A
A-7
E
F
F
F
F
TR1H1260.EPS