FujiFilm CR-IR341 Service Manual page 278

Fuji computed radiography
Table of Contents

Advertisement

1
2
3
4
C
ONTENTS
Code
Name
0985
File read error
0986
File read error
0987
DSP0 µ program load error
0A00
File open error
0A01
File format error
0A02
File setup value error
0A03
File save error
0A04
Online connection option setup error
009-051-04
05.31.99 FM2460
5
6
Significance/Occurrence Condition
An attempt to read the DspLoAD µ program file fails.
An attempt to read the DspWritE µ program file fails.
An attempt to load the DspWritE µ program fails.
An attempt is made to open a file, but it cannot be opened.
The file format is improper.
The data of the file is improper.
Cannot be saved to a file.
Both the type A online connection option and type B online
connection option are valid (option setups are inconsistent).
CR-IR341 Service Manual
Detail: Denotes the detail information format.
Probable Cause/Remedy
• No file
• HD failure
• No file
• HD failure
• There is no file in the HD.
• The file in the HDD is corrupted.
• The HDD itself is faulty, etc.
• The file in the HD format is improper.
• The file in the HDD is corrupted.
• The HDD itself is faulty, etc.
• The data described in the file in the HDD has an
out-of-range value.
• The file in the HDD is corrupted.
• The HDD itself is faulty, etc.
The HDD itself is faulty.
The type A online connection option soft-key file
and type B online connection option soft-key file are
both present in the hard disk.
MT - 19
Detail
Refer to
A-1
A-1
B
K-1
K-1
K-1
K-1
K-2
TR1H1233.EPS
MT - 19