FujiFilm CR-IR341 Service Manual page 499

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
I/O List (Sensor)
Symbol
Unit
Cassette set
SA2, 6, 10, 14
unit (A)
SA4, 8, 12, 16
SA3, 7, 11, 15
SA1, 5, 9, 13
SA17
IP
SB1
removal unit
SB2
(B)
SB3
SB4
SB5
Before-side-
SC1
SC2
positioning
conveyor (C)
Side-
SD1
positioning
conveyor (D)
SD2
SD3
BCR
After-reading
SE1
conveyor (E)
Erasure
SF1
conveyor (F)
SF2
SF3
SF4
TSW1
TSW2
LDSN1-1, 2, 3, 4, 5
Subscanning
SZ1 (SED08A)
unit (Z)
SZ2
SZ3
Models of Sensors and Parts Codes
Code No.
MS/PS
104N0043
PS
104N0044
PS
104Y0162
PS
113Y1038
PS
113Y1395
PS
128S0392
Suction sensor
128Y0241A
MS
128Y0258B
MS
009-051-04
05.31.99 FM2460
3
4
5
6
Name
Cassette IN sensor
Mammo/ST sensor
Cassette hold sensor
Cassette ejection sensor
Elevation HP sensor
Suction cup HP sensor
Cassette inlet IP sensor
Each shelf position sensor
Inch/metric sensor
Suction sensor
Before-BCR IP sensor
Convergence standby
IP sensor
Side-positioning
mechanism HP sensor
Grip release HP sensor
Side-positioning HP sensor
Barcode reader
Grip release HP sensor
After-reading conveyance
standby IP sensor
Changeover IP sensor
Load standby IP sensor
Overrun IP sensor
Safety thermostat
Safety thermostat
Unlit lamp sensor
IP leading-edge sensor
Driving-side grip release
HP sensor
Driven-side grip release
HP sensor
Model
OJ-380331-701 (5mm Type I)
OJ-380331-702 (5mm Type L) SA3, 7, 11, 15, 17, SB3, SD1
Separation type
19mm Type I
Reflective type
MPL-503(MPL)
SS-01GL111
SS-01GL2-F + mount
CR-IR341 Service Manual
7
8
9
10
Panel display
Type
When the cassette
MS
has been inserted
with the cassette
hold sensor OFF:
ON (1)
For mammo
MS
cassette: ON (1)
During cassette
PS
hold:
During cassette
PS
eject: OFF (0)
During HP: ON (1)
PS
During HP: ON (1)
PS
IP found: ON (1)
PS
When located in
the shelf position:
ON (1)
IP found: ON (1)
PS
PS (reflective
type)
During IP suction:
Suction sensor
ON (1)
IP found: ON (1)
PS
IP found: ON (1)
PS
During HP: ON (1)
PS
During HP: ON (1)
PS
IP found: ON (1)
PS
During HP: (1)
PS
IP found: ON (1)
PS
IP found: ON (1)
PS
IP found: ON (1)
PS
IP found: ON (1)
PS
(separation
type)
IP found: ON
PS (light-
(no display)
receiving type)
During HP: ON (1)
PS
During HP: ON (1)
PS
Relevant sensor symbols
SA1, 5, 9, 13, SB1, SD2, SE1, SZ2, 3
SF4
SB2, SC1, 2, SD3, SF1, 2, 3
SB4
SB5
SA2, 6, 8, 10
SA4, 8, 12, 16
Function/Remarks
Recognizes the status in pair with SA3, 7, 11,
and 15
Detects mammo and ST cassette types.
Detects the completion of cassette hold.
Detects that the cassette has been ejected.
Elevation HP for suction conveyor.
IP conveyance sensor
IP width determination
IP conveyance sensor
IP conveyance sensor
IP conveyance sensor
Variation or option
IP conveyance sensor
IP conveyance sensor
IP conveyance sensor
IP conveyance sensor
Lamp house
Fan control
IP conveyance sensor
TR1H1150.EPS
MC - 8
TR1H1003.EPS
MC - 8

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents