FujiFilm CR-IR341 Service Manual page 449

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Examples of Abnormal Images
No. 101
Tree ring-like ununiformity
• Tree ring-like ununiformity is developed.
No. 102
Horizontal streaks (1)
• Banding-like streaks are developed such that
intermittent thin horizontal streaks appear with
a gradual inclination to the main-scan direction
(from upper left to lower right).
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
<Probable Causes>
• Leaked rays of light from LAMP1 through LAMP5
may be admitted into the scanner unit.
(Improper installation of the covers, damage
on the light-tight member located inside the
cover, etc.)
• The PMT08A board is faulty.
<Check and Reference>
• Check if an error (2545: HV voltage error 2)
occurs concurrently with the abnormal image.
Note, however, that it will not be detected if the
noise is minuscule.
<Probable Cause>
• Leaked rays of light externally coming into the
machine may fall on the scanner unit. (Improper
installation of covers to the machine, especially
the lower front cover, improper installation of the
light-tight members around the light-collecting
guide assembly, etc.)
<Check and Reference>
MT - 180
FR1H1850.EPS
FR1H1851.EPS
MT - 180