FujiFilm CR-IR341 Service Manual page 495

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
2.
Common Procedures for Checks,
Replacement, and Adjustment
2.1
Turning ON/OFF High-Voltage Switch
There are two methods for turning ON/OFF the high-voltage switch: one is to directly ma-
nipulate the high-voltage switch (S1) of the SCN08A board, and the other is to use M-Utility
in a software-driven manner.
Turing ON/OFF High-Voltage Switch (Hardware Switch)
When the high-voltage switch (S1) of the SCN08A board is placed in the upper position, it is
turned ON; when it is placed in the lower position, it is turned OFF.
Scanner unit
Turning ON the High-Voltage Switch (Software Switch)
"9. HV OFF" is displayed when the high-voltage switch is in the ON position, while "9. HV
ON" is displayed when it is in the OFF position.
(1) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(2) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(3) Select "9. HV ON".
The high-voltage switch is turned ON, and "9. HV ON" displayed on the screen
switches to "9. HV OFF."
Display in
OFF status
009-051-04
05.31.99 FM2460
3
4
5
6
SCN08A board
M-Utilty screen
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV ON
10. MENU SETTING
>
CR-IR341 Service Manual
7
8
9
10
ON position OFF position
M-Utilty screen
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
Display in
8. BACKUP MEMORY
ON status
9. HV OFF
10. MENU SETTING
>
FR1H1513.EPS
FR1H1417.EPS
MC - 4
MC - 4

Advertisement