Checks after SYN08A Assembly Replacement
(1) Reinstall all the covers.
(2) Power ON the machine.
Make sure that the machine starts up normally.
(3) Generate image output, and check if there is anything abnormal with scanner control.
Uniformly expose IPs of each size at about 1 mR. Then, check the film image recorded
under SENSITIVITY from the Test Menu for scanner control failures, such as format
misalignment and shading, which may cause diagnostic problems.
If anything abnormal is found, make the following adjustments.
(4) Adjust the invalid pixels
"9.1 Adjusting Read Start Position and Read Width, [6] Fine Adjustment of Read
Width and Read Start Position"
(4) Check the SYN08A board.
" Checks"