FujiFilm CR-IR341 Service Manual page 673

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Reinstallation
(1) Attach the light-collecting mirror.
Align the right and left positioning pins of the light-collecting mirror to the respective
supports of the subscanning unit, and secure it in place with two screws.
(2) Remove the take-out screws (M3).
(3) Attach the dust cover.
While aligning the sponge located at the bottom of the dust cover against the bracket
of the light-collecting mirror, snugly fit the top sponge by hand into the scanning optics
unit.
(4) Reinstall the scanning optics unit in reverse order of removal.
(5) Install the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(6) Reinstall the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
Checks
(1) Remove the light-collecting mirror.
" Removal"
(2) Check the light-collecting mirror for any soil or scratch.
• If there is any soil:
Clean the light-collecting mirror with an ethanol-moistened lens cleaner.
" Cleaning"
• If the light-collecting mirror surface is scratched:
Generate image output and check to see that there is no ununiformity on the result-
ing image due to the scratch.
" Adjustment"
• If the ununiformity persists even after the shading/sensitivity correction:
" Replacement"
009-051-04
05.31.99 FM2460
3
4
5
6
Top view
CR-IR341 Service Manual
7
8
9
10
(1)
Support
Positioning pin
FRONT
Positioning pin
Support
MC - 182
Light-collecting
mirror
BR M4x8, x2
FR1H1543.EPS
MC - 182