10.13 Replacing Rubber Roller Located Above Subscanning Unit
(1) Remove the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(2) Remove the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
(3) Remove the light-collecting mirror
"10.4 Servicing Light-Collecting Mirror"
(4) Lift up the brace that retains the rubber roller and slide it to a position that facilitates
removal of the rubber roller.
(5) Remove the rubber roller from the subscanning unit.
(6) Remove the shaft, bearings, E-rings, and caps from the rubber roller.
(7) Install the shaft, bearings, E-rings, and caps removed at step (6) to a new rubber roller,
and reinstall the rubber roller to the subscanning unit.
Bearing
009-051-04
05.31.99 FM2460
3
4
5
6
(4)
Subscanning unit
Rubber roller
E-ring
Cap
CR-IR341 Service Manual
7
8
9
10
Rubber roller
Spring
Cap
E-ring
Bearing
Shaft
MC - 199
Brace
FR1H1528.EPS
MC - 199
FR1H1526.EPS