(6) Detailed Sequence of Driven Shaft Self-Diagnostics
Fig. 5.9.6 Driven Shaft Self-Diagnostics
Preprocessing: Fig. 5.9.1
START
Speed data setting
Driven shaft grip release driving
DRIVE (MZ1/ST speed ON)
TZ35 STRAT
Timer START
Is SZ3 opened within TZ35?
N
Y
Driven shaft grip release stop
DRIVE (MZ3/OFF) (hard)
Timer STOP
Timer>TZ37 or Timer< TZ36
N
Y
Rotation-speed
correction
Sequence flow chart
See 5.9 (8).
Speed data update
DELAY TZ51
Speed data setting
END
Table 5.9.6 Operation Parameters
Symbol
Driven shaft grip release timeout (self-diagnostics)
TZ35
Lower limit for driven shaft grip release spec (self-diagnostics)
TZ36
Upper limit for driven shaft grip release spec (self-diagnostics)
TZ37
Wait for initialization grip operation
TZ51
009-051-04
05.31.99 FM2460
3
4
5
6
The previous latest data
(corrected) in COMS is used.
Driven shaft grip release stop
DRIVE(MZ3/OFF) (soft)
03CB
Description
CR-IR341 Service Manual
7
8
9
10
Level 0 error
Driven shaft grip release
self-diagnostics error
FR1H0435.EPS
Value
5 sec
1.7 sec
2.0 sec
0.5 sec
TR1H0085.EPS
MD - 148