FujiFilm CR-IR341 Service Manual page 223

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
6.7
Error Handling: IDT Line Error
(1) Overall/General Sequence
This error handling is performed when IDT line error occurs in the CSL/IDT information
check sequence.
Corresponding sequence
Fig. 2.6.6 CSL/IDT Information Check
At the time when
the error IP is
returned to the
cassette, an
elevation
permission
is issued to
start the processing
for the next IP in the
queue. At the same time,
removal of the cassette
for the error IP is notified
to the user.
Cassette ejection
Same as in the routine
described in Fig. 2.11.1.
If hold release error
occurs during cassette
ejection, degeneration
takes place for the
relevant shelf.
009-051-04
05.31.99 FM2460
3
4
5
6
Fig. 6.7.1 Error Handling: IDT Line Error
Preprocessing: Fig. 2.6.6
START
IDT line established
Process stopped
Cassette removed
N
Y
N
Y
END
CR-IR341 Service Manual
7
8
9
10
No matter whether there is any IP in the
machine or there is any unprocessed
cassette, it is placed in a queue until
Error message output
an elevation permission.
Stop ARARM
/UNLOAD CASS.
Alarm disabled
Suction cup movement (HP
DRIVE (MB1/PB14)
This elevation permission allows the
Reverse conveyance
process to start.
load
Fig.6.3.2
Elevation permission
is issued
Error message output
Stop ARARM/UNLOAD CASS.
Alarm disabled
Cassette hold release
Fig.2.2.2
The cassette removable condition is
LEDT2 is turned on
indicated.
Cassette removal
Operation performed by the operator
Is SA1 opened?
DELAY TA13
Instant return of cassette is supported.
Is SA2 opened?
LEDT2 is turned off
The ready condition is indicated.
LEDT1 is turned on
Return to normal routine
Another IP in the machine proceeds
to the load standby condition.
The unprocessed cassette enters the
feed standby condition.
load standby)
Suction cup load preparation
If there is another IP in the machine,
up-down load
load.
If there is any unprocessed cassette,
up-down feed
feed.
IDT line error