FujiFilm CR-IR341 Service Manual page 602

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
Reinstallation
CAUTION
When removing and installing the units, exercise care not to damage the cables.
NOTE
Make sure that the scanner unit has been pulled all the way out. If not, MD3 will interfere with
the machine frame when the side-positioning conveyor is removed.
(1) Reinstall the side-positioning conveyor.
While pressing the bracket of the side-positioning conveyor against the block section
of the subscanning unit and the reference side, secure the side-positioning conveyor in
place with five screws.
(2) Connect the connectors (CND1, BCRCN1, CNMD1-3).
Block section
(3) Push in the scanner unit.
"7. Pulling Out and Pushing in Scanner Unit"
009-051-04
05.31.99 FM2460
3
4
5
6
Top view
(1)
Block section of subscanning unit
Reference
side
Bracket
Side-positioning conveyor
Bracket
(2)
CND1
(2) BCRCN1
CR-IR341 Service Manual
7
8
9
10
Bracket
(2)
CNMD3
CNMD3
(1)
(1)
Side-positioning conveyor
BR M4x8, x5
MC - 111
MD3
(2)
CNMD1, CNMD2
FR1H1773.EPS
MC - 111

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus