ONTENTS
2.3.2 Format of Abort Code
If any trouble occurs on the machine with an abort message displayed on the machine's display panel, take remedial action in accordance with
the following procedures.
Example of Abort Message
Abort message
CDM:CdmInit:Could not allocate
memory for config data.
matrix_select:Undefined event Matrix
=AAAAAAAA EVENT=BBBBBBBB
"Memory overflow."
Remedy
(1) Write down the top three liens of the message displayed, and notify them to the Service Headquarters.
(2) Reset the machine.
009-051-04
05.31.99 FM2460
5
6
Abort code
Significance
02120000
Failure to reserve the machine configuration area
02010001
Undefined event received
05050001
Failure to reserve memory
CR-IR341 Service Manual
Probable cause
Insufficient main memory
Illegal function call
Insufficient main memory
TR1H1302.EPS