6.9
Checking and Replacing Shock Absorber-Attached Guide
of Before-Side-Positioning Conveyor
Check and Replacement
Checking and replacing shock absorbers located above before-side-
positioning conveyor and under cleaning roller
(1) Remove the before-side-positioning conveyor.
"6.1 Removing and Reinstalling Before-Side-Positioning Conveyor"
(2) Remove the cover of the before-side-positioning conveyor.
(3) Loosen the single screw that retains the bracket.
(4) Remove the shock absorber-attached guide plate (A).
(5) Remove the cleaning roller assembly.
(6) Remove the shock absorber-attached guide plate (B).
(7) Check if the shock absorber is not peeled or frayed.
If it is peeled or frayed, replace the whole guide to which the shock absorber is at-
tached.
(5)
Cleaning roller assembly
009-051-04
05.31.99 FM2460
3
4
5
6
(3)
Loosen screw
BR M4x8
(3)
Bracket
Before-side-positioning
conveyor
BR M4x8
CR-IR341 Service Manual
7
8
9
10
(2)
Cover
BR M4x8, x4
MC - 100
(4), (7)
Shock absorber-attached
guide plate (A)
Sems M3x6, x4
(6),(7)
Shock absorber-attached
guide plate (B)
Sems M3x6, x2
FR1H1549.EPS
MC - 100